Publications

Zhang, LF; Huang, CP; Wu, TX; Zhang, FZ; Tong, QX (2011). Laboratory Calibration of a Field Imaging Spectrometer System. SENSORS, 11(3), 2408-2425.

Abstract
A new Field Imaging Spectrometer System (FISS) based on a cooling area CCD was developed. This paper describes the imaging principle, structural design, and main parameters of the FISS sensor. The FISS was spectrally calibrated with a double grating monochromator to determine the center wavelength and FWHM of each band. Calibration results showed that the spectral range of the FISS system is 437-902 nm, the number of channels is 344 and the spectral resolution of each channel is better than 5 nm. An integrating sphere was used to achieve absolute radiometric calibration of the FISS with less than 5% calibration error for each band. There are 215 channels with signal to noise ratios (SNRs) greater than 500 (62.5% of the bands). The results demonstrated that the FISS has achieved high performance that assures the feasibility of its practical use in various fields.

DOI:
1424-8220

ISSN:
10.3390/s110302408